Radiation dose and image quality for adult interventional cardiology in Chile: A national survey

  • C. Ubeda
  • , E. Vano
  • , P. Miranda
  • , F. Leyton
  • , E. Valenzuela
  • , C. Oyarzun

Producción científica: Contribución a una revistaArtículorevisión exhaustiva

8 Citas (Scopus)

Resumen

The aim of this work was to investigate the differences in dose settings and image quality among 10 X-ray systems used for interventional cardiology in Chile. Entrance surface air kerma (ESAK) was measured on a phantom of 20 cm thickness of polymethyl methacrylate slabs. Image quality was evaluated using DICOM images of a test object Leeds TOR 18-FG for cine mode acquisition, through the numerical parameters signal-to-noise ratio (SNR), high-contrast spatial resolution (HCSR) and figure of merit. ESAK rate values for fluoroscopy modes ranged between 7.1 and 121.7 mGy min -1. For cine mode, ESAK values per frame ranged from 63 to 400 μGy fr -1. SNR and HCSR parameters for cine mode varied from 4.8 to 8.6 and 0.4 to 10, respectively. FOM values resulted from 6.9 to 64.5 among the different X-ray systems. Results show important differences between systems and point out the need to launch an optimisation programme.