The aim of this work was to investigate the differences in dose settings and image quality among 10 X-ray systems used for interventional cardiology in Chile. Entrance surface air kerma (ESAK) was measured on a phantom of 20 cm thickness of polymethyl methacrylate slabs. Image quality was evaluated using DICOM images of a test object Leeds TOR 18-FG for cine mode acquisition, through the numerical parameters signal-to-noise ratio (SNR), high-contrast spatial resolution (HCSR) and figure of merit. ESAK rate values for fluoroscopy modes ranged between 7.1 and 121.7 mGy min -1. For cine mode, ESAK values per frame ranged from 63 to 400 μGy fr -1. SNR and HCSR parameters for cine mode varied from 4.8 to 8.6 and 0.4 to 10, respectively. FOM values resulted from 6.9 to 64.5 among the different X-ray systems. Results show important differences between systems and point out the need to launch an optimisation programme.
| Idioma original | Inglés |
|---|---|
| Número de artículo | ncr276 |
| Páginas (desde-hasta) | 90-93 |
| Número de páginas | 4 |
| Publicación | Radiation Protection Dosimetry |
| Volumen | 147 |
| N.º | 1-2 |
| DOI | |
| Estado | Publicada - sep. 2011 |