Paediatric interventional cardiology: Flat detector versus image intensifier using a test object

  • E. Vano
  • , C. Ubeda
  • , L. C. Martinez
  • , F. Leyton
  • , P. Miranda
  • Complutense University
  • Hospital Universitario 12 de Octubre
  • Ministerio de Salud de Chile
  • Luis Calvo Mackenna Hospital

Research output: Contribution to journalArticlepeer-review

15 Scopus citations

Abstract

Entrance surface air kerma (ESAK) values and image quality parameters were measured and compared for two biplane angiography x-ray systems dedicated to paediatric interventional cardiology, one equipped with image intensifiers (II) and the other one with dynamic flat detectors (FDs). Polymethyl methacrylate phantoms of different thicknesses, ranging from 8 to 16 cm, and a Leeds TOR 18-FG test object were used. The parameters of the image quality evaluated were noise, signal-difference-to-noise ratio (SdNR), high contrast spatial resolution (HCSR) and three figures of merit combining entrance doses and signal-to-noise ratios orHCSR. The comparisons showed a better behaviour of the II-based system in the low contrast region over the whole interval of thicknesses. The FD-based system showed a better performance in HCSR. The FD system evaluated would need around two times more dose than the II system evaluated to reach a given value of SdNR; moreover, a better spatial resolution was measured (and perceived in conventional monitors) for the system equipped with flat detectors. According to the results of this paper, the use of dynamic FD systems does not lead to an automatic reduction in ESAK or to an automatic improvement in image quality by comparison with II systems. Any improvement also depends on the setting of the x-ray systems and it should still be possible to refine these settings for some of the dynamic FDs used in paediatric cardiology.

Original languageEnglish
Pages (from-to)7287-7297
Number of pages11
JournalPhysics in Medicine and Biology
Volume55
Issue number23
DOIs
StatePublished - 7 Dec 2010

Fingerprint

Dive into the research topics of 'Paediatric interventional cardiology: Flat detector versus image intensifier using a test object'. Together they form a unique fingerprint.

Cite this