Abstract
Our objective was to obtain images of a predictable level of quality using an intraoral X-ray system with digital imaging, avoiding patient overexposure. A polymethylmethacrylate (PMMA) physical test phantom was imaged at different exposure times and at various PMMA thicknesses using a dental imaging coupled charge device. Two identical regions of interest (ROIs) were chosen in every image file, and quality was numerically evaluated by measuring high-contrast spatial resolutions, low-contrast thresholds and signal-to-noise ratios. In addition, three practitioners proposed personal quality scores by image inspection. Numerical contents in the ROIs, related to the image quality, were plotted against exposure time. From here, a simple expression linking the exposure time with the thickness to obtain images of comparable quality was deduced. As a result, the optimum exposure time for imaging with a predictable level of quality can be inferred. The potential effect could imply savings above 1000 man Sv, roughly 20 % of the collective dose due to dental imaging, over a population of 1540 millions.
| Original language | English |
|---|---|
| Article number | ncr470 |
| Pages (from-to) | 95-101 |
| Number of pages | 7 |
| Journal | Radiation Protection Dosimetry |
| Volume | 151 |
| Issue number | 1 |
| DOIs | |
| State | Published - Aug 2012 |
UN SDGs
This output contributes to the following UN Sustainable Development Goals (SDGs)
-
SDG 3 Good Health and Well-being
Fingerprint
Dive into the research topics of 'Evaluating phantom image quality parameters to optimise patient radiation dose in dental digital radiology'. Together they form a unique fingerprint.Cite this
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver