Comparison of image qualityamong three X-ray systems for chest radiography: First step in optimisation

  • Ernesto Torres Galdames Regional Hospital
  • Koninklijke Philips N.V.

Research output: Contribution to journal › Article › peer-review

5 Scopus citations

Abstract

The aim of this study was to compare the performance of three digital X-ray systems [one flat-panel (DR) and two computed radiography (CR)] for chest radiography in terms of the entrance surface air kerma (ESAK) delivered to a polymethyl methacrylate phantom of 20 cm (equivalent to an adult patient) and image quality through of numerical evaluations using a test object (TO). The tube charge applied was ranged from 0.6 to 32 mAs, to a fixed tension of 125 kVp. The DR system presented the highest mean values of ESAK (615.9 μGy) along with the highest signal-to-noise ratio values, whereas CR systems showed a better high-contrast spatial resolution. Differences were statistically significant in both cases regarding the tube charge used. Thus, this parameter should be mainly considered to optimise the radiological protection through exposure settings selected. This survey represents the first effort to achieve optimisation in digital radiology for Chile.

Original languageEnglish
Pages (from-to)386-391
Number of pages6
JournalRadiation Protection Dosimetry
Volume165
Issue number1-4
DOIs
StatePublished - 1 Jul 2015

UN SDGs

This output contributes to the following UN Sustainable Development Goals (SDGs)

  1. SDG 3 - Good Health and Well-being
    SDG 3 Good Health and Well-being

Fingerprint

Dive into the research topics of 'Comparison of image qualityamong three X-ray systems for chest radiography: First step in optimisation'. Together they form a unique fingerprint.

Cite this