Abstract
This paper presents the results of the first characterization and experimental measurements of scatter dose at cardiologist's eyes for the only X-ray system that performs all paediatric Interventional cardiology procedures in Costa Rica. Entrance surface air kerma (ESAK) and the scatter dose values were measured on phantoms of 4-20 cm thicknesses of polymethyl methacrylate slabs. Image quality was evaluated using DICOM images of a test object Leeds TOR 18-FG, through the numerical parameters signal-to-noise ratio (SNR), high-contrast spatial resolution (HCSR) and figure of merit (FOM). When PMMA thickness is increased and during a move from low fluoroscopy to cine modes, ESAK, SNR, HCSR and FOM values range from 0.44 to 391.0 μGy fr-1; 2.8 to 14.89; 3.17 to 15.54 and 0.51 to 79.94, respectively. The highest scattered dose rates recorded during the simulations were 1.79 and 8.04 mSv h-1 for the high fluoroscopy and cine modes, respectively.
| Original language | English |
|---|---|
| Pages (from-to) | 450-455 |
| Number of pages | 6 |
| Journal | Radiation Protection Dosimetry |
| Volume | 176 |
| Issue number | 4 |
| DOIs | |
| State | Published - 1 Nov 2017 |
UN SDGs
This output contributes to the following UN Sustainable Development Goals (SDGs)
-
SDG 3 Good Health and Well-being
Fingerprint
Dive into the research topics of 'Characterization and experimental measurements of scatter dose at cardiologist's eyes during paediatric interventional cardiology procedures in Costa Rica'. Together they form a unique fingerprint.Cite this
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver