Biplane interventional pediatric system with cone-beam CT: Dose and image quality characterization for the default protocols

  • Eva Corredoira
  • , Eliseo Vañó
  • , Luis Alejo
  • , Carlos Ubeda
  • , Federico Gutiérrez-Larraya
  • , Julia Garayoa
  • Hospital Universitario La Paz
  • Complutense University
  • Hospital Universitario Fundación Jiménez Díaz

Research output: Contribution to journalArticlepeer-review

6 Scopus citations

Abstract

The aim of this study was to assess image quality and radiation dose of a biplane angiographic system with cone-beam CT (CBCT) capability tuned for pediatric cardiac procedures. The results of this study can be used to explore dose reduction techniques. For pulsed fluoroscopy and cine modes, polymethyl methacrylate phantoms of various thicknesses and a Leeds TOR 18-FG test object were employed. Various fields of view (FOV) were selected. For CBCT, the study employed head and body dose phantoms, Catphan 504, and an anthropomorphic cardiology phantom. The study also compared two 3D rotational angiography protocols. The entrance surface air kerma per frame increases by a factor of 3-12 when comparing cine and fluoroscopy frames. The biggest difference in the signal-to-noise ratio between fluoroscopy and cine modes occurs at FOV 32 cm because fluoroscopy is acquired at a 1440 x 1440 pixel matrix size and in unbinned mode, whereas cine is acquired at 720 x 720 pixels and in binned mode. The high-contrast spatial resolution of cine is better than that of fluoroscopy, except for FOV 32 cm, because fluoroscopy mode with 32 cm FOV is unbinned. Acquiring CBCT series with a 16 cm head phantom using the standard dose protocol results in a threefold dose increase compared with the low-dose protocol. Although the amount of noise present in the images acquired with the low-dose protocol is much higher than that obtained with the standard mode, the images present better spatial resolution. A 1mm diameter rod with 250 Hounsfield units can be distinguished in reconstructed images with an 8mm slice width. Pediatric-specific protocols provide lower doses while maintaining sufficient image quality. The system offers a novel 3D imaging mode. The acquisition of CBCT images results in increased doses administered to the patients, but also provides further diagnostic information contained in the volumetric images. The assessed CBCT protocols provide images that are noisy, but with very good spatial resolution.

Original languageEnglish
Pages (from-to)357-376
Number of pages20
JournalJournal of Applied Clinical Medical Physics
Volume17
Issue number4
DOIs
StatePublished - 2016

Keywords

  • Cone-beam CT
  • Dosimetry
  • Image quality
  • Pediatric cardiac catheterization
  • Rotational angiography

Fingerprint

Dive into the research topics of 'Biplane interventional pediatric system with cone-beam CT: Dose and image quality characterization for the default protocols'. Together they form a unique fingerprint.

Cite this